Main advantages
The design of the respiratory system provides the primary convenience of sterilization (a collapsible design, the maintenance of which does not cause difficulties for personnel).
The battery is designed for 60 minutes of operation (autonomous charging, automatic transition to backup power when instability is detected or there is no power supply).
There is a manual ventilation mode (pressure relief valve has a wide range of settings).
The ability to connect a backup oxygen source (the transition is carried out smoothly, without stopping the workflow).
High informativeness of the display, display of all parameters in a graphic form.
Ample opportunities for modernization (mount for vaporizers, multi-gas monitor, capnograph, vaporizers for different types of anesthesia, other equipment and accessories).
NDA Orpheus supports the following ventilation modes: forced (controlled by volume or pressure), synchronized (controlled by volume), pressure control mode with monitoring and limitation of peak pressure, independent, adjustable trigger based on pressure or flow, manual.
Specifications
Minute ventilation range: 0-35 l/min.
Tidal volume control range: 20–1600 ml.
The ratio of the duration of inhalation and exhalation: 5:1-1:10.
PEEP control limits: 0-30 cm water column.
Adjustable ventilation frequency: 2-100 per minute.
Limits for setting the duration of the inspiratory pause: 0 - 50%.
Inspiratory pressure setting limits: 1–100 cm of water column.
Threshold of maximum inspiratory pressure: 1–100 cm of water column.
Oxygen supply regulation range: 0.2–10 l/min.
Nitrous oxide supply control range: 0.2–10 l/min.
Air supply regulation range: 1–20 l/min.
Adsorber volume: 1.6 l.
Automatic protection against the supply of hypoxic breathing mixture.
Overpressure relief valve.
Emergency air supply valve.
Time of continuous operation of the built-in backup power supply: not less than an hour.
Ventilation modes
Forced ventilation controlled by volume CMC-VC.
Forced ventilation controlled by CMC-PC pressure.
Volume controlled ventilation with additional CMV(PV) peak pressure limitation.
Pressure controlled ventilation with CMV(PC) peak pressure limitation.
Volume-controlled synchronized intermittent ventilation SIMV-VC.
Synchronized intermittent pressure controlled ventilation SIMV-PC.
Spontaneous breathing in the airways with inspiratory support by CPAP + PS pressure.
Manual ventilation.
The work of the trigger on the thread.
Pressure trigger operation.